What is a Spectrophotometer?
How does a Spectrophotometer for color measurement work?
Measurement Geometry: d/8° vs. 45°/0°
SCI and SCE: What do they mean?
What does a Spectrophotometer measure?
How to use a Spectrophotometer Step by Step?
Measurement Apertures: Choosing the Right Size
Applications of Spectrophotometers in Industry
How do the measurement instruments in our product range differ?
How to read a Spectrophotometer Specification?
Frequently Asked Questions
The term “spectrophotometer” appears across many industries – from laboratory chemistry to quality control in the coatings industry. The challenge is that the same term can refer to instruments that operate according to completely different principles and are used for different purposes.
This article focuses on spectrophotometers for color measurement – instruments that measure color using spectral reflectance and are used in industrial production.
What is a Spectrophotometer?
A spectrophotometer is a measuring instrument that determines how a sample absorbs or reflects light depending on its wavelength. The result is a spectral reflectance curve – a unique “color signature” of the sample, based on which color coordinates are calculated in colorimetric color spaces such as CIELab, CIELCh, XYZ, and others.
Simply put, a spectrophotometer answers the question: “How would this color appear to an observer’s eye under a specific illuminant?” – and expresses the answer in numerical values.
Color Measurement Spectrophotometers vs. Laboratory Spectrophotometers: An Important Distinction
On the internet, the same term is used for two categories of instruments:
| Feature | UV-VIS Spectrophotometer (Laboratory Instrument) | Color Measurement Spectrophotometer (3Color®) |
| Operating Principle | UV and VIS absorption/transmittance | Spectral reflectance in the VIS range (360–780 nm) |
| What Is Measured? | Concentration of substances in a solution | Color of a surface or material |
| Result | Absorbance, concentration (mg/L, ppb) | CIELab, CIELCh, ΔE, reflectance |
| Typical Applications | Chemistry, biology, pharmaceuticals | Paints and coatings, plastics, textiles, printing industry |
| Sample | Solution in a cuvette | Solid surface or film |
If you are looking for an instrument for color control in industrial production, this article is for you. If, on the other hand, you need a spectrophotometer for chemical analysis (DNA, enzymes, concentrations), you are looking at a completely different category of instrument.
How does a Spectrophotometer for color measurement work?
The operating principle is based on measuring spectral reflectance: the instrument illuminates the sample and measures how much light is reflected from its surface at each wavelength within the visible range (360–780 nm), in 10 nm increments.
Step by Step: What happens inside the instrument?
- The light source illuminates the sample with a standardized beam of light.
- An integrating sphere (d/8° geometry) or a 45°/0° system evenly distributes or directs the light onto the sample.
- A diffraction grating separates the reflected light into individual wavelengths.
- A CMOS sensor measures the reflectance intensity at each wavelength from 360 to 780 nm in 10 nm increments.
- The software processes the reflectance curve into color coordinates (L*, a*, b*) and calculates the ΔE deviation compared with the reference sample.
The entire measurement takes approximately 1 second.
Calibration Before Measurement
Before each series of measurements, the spectrophotometer performs a calibration using a white and a black reference tile. The white tile (ceramic) defines approximately 100% reflectance, while the black tile establishes the zero point (light trap). Without calibration, the results may be affected by thermal drift and aging of the light source. 3Color® spectrophotometers (e.g., the SFXn Series) use intelligent automatic calibration. The instrument detects the conditions automatically and prompts the user to perform calibration when a temperature threshold is exceeded or a specified number of measurements has been reached.
Measurement Geometry: d/8° vs. 45°/0°
The measurement geometry describes the arrangement of the light source and detector relative to the sample. It is the most important parameter when selecting a spectrophotometer. Choosing the wrong geometry can result in measurements that are not comparable with those from the customer’s or supplier’s laboratory.
| Parameter | d/8° (Spherical Geometry) | 45°/0° (Circumferential Geometry) |
| Illumination | Diffuse (integrating sphere) | At 45° to the surface |
| SCI/SCE | Yes, two measurement modes | SCE only (gloss excluded) |
| Compliance Standards | ISO 7724, ASTM E1164, CIE No.15 | ASTM E1349, ISO 5-4 |
| Applications | Plastics, coatings, cosmetics, textiles | Printing industry, printing inks, automotive |
| 3Color® Models | SFXn, SFXne, SV-400, SV-300, Neo 9001 | SFXnc, SFXnc-g, SV-400c |
SCI and SCE: What do they mean?
SCI (Specular Component Included): Measurement with the specular component included. This measures color independently of surface gloss. It is ideal for comparisons between laboratories when samples have different surface structures.
SCE (Specular Component Excluded): Measurement with the specular component excluded. It corresponds to the visual appearance: a matte sample and a glossy sample with the same pigment color will produce different measurement results – just as the human eye perceives them.
Spherical spectrophotometers (d/8°) perform both measurement modes in a single measurement. This is an important advantage over the 45°/0° geometry, which measures exclusively in SCE mode.
What does a Spectrophotometer measure?
A color measurement spectrophotometer measures spectral reflectance (0–200%) across the 360–780 nm range. Based on this data, it calculates dozens of colorimetric and optical parameters:
Color Coordinates
- CIELab (L, a, b*):** The most widely used industry standard
- CIELCh (L, C, h°):** Color represented in terms of lightness, saturation, and hue
- CIEXYZ: CIE tristimulus values, the basis of all color spaces
- HunterLab: An older color space still used in some industries, such as paper and plastics
- Yxy: Luminance and chromaticity
- RGB: Channel values for digital applications
Color Difference (ΔE)
ΔE (Delta E) is a measure of the color difference between a sample and a reference standard. The lower the ΔE value, the more closely the colors match. In quality control, acceptable tolerances are defined:
| Formula | ΔE Range / Rating | Typical Application |
| ΔE*ab (CIE 1976) | < 1: no perceptible difference | General industry standard |
| ΔE*00 (CIEDE 2000) | < 0.5: excellent match | Paints, coatings, cosmetics. Highest accuracy |
| ΔE*94 | < 1: acceptable in production | Textiles, printing industry |
| ΔE*CMC | < 1: no visible difference | Textile industry (ISO 105-J03) |
Learn more about color difference in our dedicated article on color measurement.
Colorimetric Indices
- Whiteness Index: ASTM E313, CIE, ISO 2470, Hunter – evaluation of the whiteness of paper, plastics, and textiles
- Yellowness Index: ASTM D1925, ASTM E313 – evaluation of material aging and plastic quality
- Color Strength: Quality control of pigments and color concentrates
- Blackness (My, dM): Depth of black in the printing and textile industries
- Metamerism: Evaluation of whether two colors appear identical under different illuminants
- CMYK Optical Density: For the printing and offset printing industries
- Hiding Power: Opacity of paint and coating layers
Optical Properties
- Spectral Reflectance [%] at each wavelength from 360–780 nm
- Transmittance (T Mode) for transparent samples: films, glass, liquids
- Munsell: Conversion to the traditional Munsell color system
How to use a Spectrophotometer Step by Step?
Measurement Preparation:
- Turn on the instrument and wait at least 5 minutes for the light source to reach thermal stability.
- Perform calibration using the white and black reference tiles included with the instrument.
- Select the observer angle: 2° (small samples, laboratory applications) or 10° (production, visual quality control).
- Select the standard illuminant: D65 (standard daylight, most commonly used), D50 (graphics, printing industry), A (incandescent light), F2/F11 (fluorescent lamps).
- Select the SCI or SCE mode as required.
- Select the ΔE formula according to customer requirements or the applicable industry standard.
Reference Sample Measurement:
- Place the reference sample (standard) over the measurement aperture. The sample must completely cover the aperture.
- Press the measurement button. The result appears within approximately 1 second.
- Save the reference sample in the instrument’s memory or in the software’s sample library.
Sample Measurement:
- Place the sample (production batch) in the same position on the instrument as the reference sample.
- Perform the measurement. The instrument automatically compares the sample with the reference sample and displays the ΔE value.
- If ΔE < tolerance: batch accepted. If ΔE > tolerance: deviation flagged for further analysis.
For reliable results, the measurement should be repeated 3–5 times and the average value calculated, especially for inhomogeneous samples.
Common Measurement Errors
| Error | Consequence and Solution |
| No calibration before measurement | Measurement drift and incorrect ΔE values. Calibrate each time the instrument is switched on. |
| Sample does not completely cover the aperture | The sample edge is measured instead of the center. Use a suitable aperture. |
| Inconsistent sample pressure | Different air layer thicknesses. Apply constant and reproducible pressure. |
| Changing the illuminant between measurements | Results are not comparable. Always use the same illuminant and observer. |
| Measurement on loose fabric | Variable sample pressure. Use a backing or measurement frame. |
| Dirty measurement aperture | Distorted results. Clean the aperture with a lint-free cloth. |
Measurement Apertures: Choosing the Right Size
The 3Color® SFXn Series spectrophotometers feature interchangeable measurement apertures. Choosing the right aperture size is essential for reliable measurements:
| Measurement Aperture | When to Use | Examples of Materials |
| Φ11 mm (Standard) | Large, homogeneous samples | Plastic sheets, textiles |
| Φ6 mm | Small, inhomogeneous samples | Cosmetic samples, small labels |
| Φ3 mm (Small) | Very small details | Fine fabric patterns, automotive components |
| 1 × 3 mm (SFXne) | Narrow samples, yarn | Threads, profile edges, narrow paint strips |
| Φ5 mm (SFXne) | Medium-sized samples | Printing inks, small ceramic parts |
General Rule: The measurement aperture should be selected as large as possible for the given sample. A measurement area that is too small amplifies texture inhomogeneities and increases the variation in measurement results.
Applications of Spectrophotometers in Industry
Paints and Coatings
Color control of paint and coating finishes – from pigment formulation to final production batch approval. The spectrophotometer enables formulation (Computer Color Matching), allowing a pigment formulation to be calculated based on the measurement of a reference sample. Key parameters: ΔE*00, Metamerism Index, Color Strength, Hiding Power. Recommended geometry: d/8° (SFXn) for formulation and laboratory applications, and 45°/0° (SFXnc) for visual assessment of the coating appearance.
Plastics
Evaluation of the color of granules, injection-molded parts, and finished components – particularly important in the automotive and packaging industries. The spectrophotometer detects color deviations caused by different pigment batches, changes in injection molding temperature, or UV aging of the material. A typical tolerance for visible components is: ΔE*ab < 0.5. Recommended geometry: d/8° SCI.
Textiles and Apparel
Control of color matching for fabric, yarn, and thread batches against the reference sample. In the textile industry, ΔECMC or ΔE94 formulas and metamerism reports are used to evaluate color under standard illuminants D65, A, and F11. The small 1 × 3 mm aperture (SFXne) enables measurement of individual threads or narrow fabric strips.
Printing Industry
Verification of printing inks against the ICC profile. The 45°/0° geometry (SFXnc) corresponds to the visual appearance perceived by the observer. Measuring CMYK optical density and CIELab coordinates enables a ΔE < 2 to be maintained between different print runs. Standards: ISO 12647-2 (offset printing), ISO 12647-3 (newspaper printing).
Cosmetics and Pharmaceuticals
Evaluation of the color of creams, powders, lipsticks, and active ingredients. Both the color of the product itself and its behavior under different lighting conditions (metamerism) are important. The spectrophotometer enables color standards to be defined and maintained across production batches.
Ceramics, Glass, and Enamel
Color control of glazes, enamels, and glassware. Transmittance measurement (T mode) is used for transparent or translucent samples. The d/8° geometry eliminates the influence of gloss effects on the color measurement result.
How do the measurement instruments in our product range differ?
This is one of the most frequently asked questions. The short answer is: a spectrophotometer measures the spectral reflectance curve, while a colorimeter measures only the tristimulus values. This difference determines the instrument’s range of functions and measurement accuracy.
| Parameter | Spectrophotometer | Colorimeter |
| Measures | Reflectance curve (multiple measurement points) | RGB channels → XYZ → Lab |
| ΔE Accuracy | ΔE*ab ≤ 0.01 (repeatability) | ΔE*ab ≤ 0.08–0.15 |
| Metamerism Analysis | Yes, complete spectral data | No, no spectral data |
| Change Illuminant Without New Measurement | Yes, conversion based on the reflectance curve | No, a new measurement is required |
| Color Formulation | Yes | No |
| Price | Higher | Lower |
| When to Choose? | Laboratory, color formulation, quality control | Production control with a limited budget |
For a detailed comparison between colorimeters and spectrophotometers, see the article: Colorimeter vs. Spectrophotometer.
How to read a Spectrophotometer Specification?
When purchasing a spectrophotometer, it is important to understand the key parameters listed in the product specifications:
| Parameter | Meaning and What to Look For |
| Spectral Range | 360–780 nm: full visible range. Models with 400–700 nm may omit important fluorescence data. |
| Spectral Interval | 10 nm: industry standard. A smaller interval provides more measurement points and a more precise reflectance curve. |
| Repeatability ΔE*ab | ≤ 0.01: top-level market performance. Measures the variation when the same sample is measured repeatedly. |
| Inter-Instrument Agreement | ≤ 0.2 ΔE*ab: difference between individual instruments. Critical for connected workflows using multiple devices. |
| Integrating Sphere Diameter | Φ40 mm: spherical standard. A larger sphere provides greater illumination uniformity. |
| Light Source Lifetime | 10,000,000 measurements (SFXn): critical for total cost of ownership (TCO). Don’t focus only on the purchase price. |
| Connectivity | USB + Bluetooth: high flexibility. Bluetooth for use at the production line and USB in the laboratory. |
| Supported Illuminants | The more, the better. At minimum: D65, D50, A, F2, F11. |
Frequently Asked Questions
What Is a Spectrophotometer?
A spectrophotometer is a measuring instrument that measures the spectral reflectance of a sample in the visible range (360–780 nm). Based on the spectral data, it calculates color coordinates (CIELab, CIELCh), color difference ΔE, and colorimetric indices such as whiteness, yellowness, and color strength. In industry, it is used for objective, numerical color control.
How Does a Spectrophotometer for Color Measurement Work?
The spectrophotometer illuminates the sample with a standardized light source (LED + xenon lamp) and then records how much light is reflected from the surface at each wavelength from 360 to 780 nm. A diffraction grating separates the reflected light into its spectral components. The software processes the results and calculates color coordinates and the ΔE value compared with the reference sample.
What Does a Spectrophotometer Measure?
A color measurement spectrophotometer measures spectral reflectance [%] at each wavelength, color coordinates (L*, a*, b*, C*, h°), color differences ΔE (ΔEab, ΔE00, ΔECMC, ΔE94), whiteness, yellowness, and blackness indices, color strength, CMYK optical density, and metamerism.
What Is the Difference Between a Spectrophotometer and a Colorimeter?
A spectrophotometer measures the complete reflectance curve (40+ measurement points), enabling metamerism analysis, changing the illuminant without taking a new measurement, and color formulation. A colorimeter measures only XYZ tristimulus values using RGB filters. It is simpler and less expensive, but does not provide spectral data.
Recommendation: A spectrophotometer is recommended for laboratories and demanding quality control applications, while a colorimeter is suitable for basic color checks in production.
What Is Delta E (ΔE)?
Delta E (ΔE) is a numerical value that describes the color difference between a sample and a reference sample. The lower the ΔE value, the more similar the colors are. In general, ΔE < 1 is not perceptible to the naked eye, ΔE 1–2 is barely visible, and ΔE > 3 is clearly perceptible. In the coatings industry, a tolerance of ΔE*00 < 0.5 is often used.
Wie verwendet man ein Spektralphotometer?
- Turn on the instrument and wait 5 minutes for stabilization.
- Perform calibration using the white and black reference tiles.
- Select the observer (2° or 10°), illuminant (D65, D50, A), and ΔE formula.
- Measure the reference sample and save it in the instrument’s memory.
- Measure the production samples. The instrument automatically displays the ΔE value.
- Accept batches with a ΔE value within the tolerance and reject or further analyze batches with higher values.